Publication: Dissolution-guided wetting for microarray and microfluidic devices.

The widespread use of lab-on-a-chip devices containing microcavities in the form of microwells, traps, dead ends, and corners, is often precluded by the trapping of air bubbles. Allbritton and colleagues present a straightforward and simple approach to mitigating bubble formation that can be applied to the CellRaft™ Arrays that are the cornerstone of the CellRaft® Technology. Air bubbles are eliminated in a two-step process, whereby microarrays are first hydrophilized by plasma treatment, and subsequently coated with a monosaccharide such as D-glucose or D-sorbitol. Following this treatment, microwell arrays can be stored for as long as 6 months in air, and complete rewetting of the microwells is demonstrated by the dissolution of the monosaccharide with an aqueous solution.

Related Resources

In this study the CERO 3D Incubator & Bioreactor was used for the generation of tumor spheroids from human colon cancer cell line HT29-MTX-E12. The...

The paper discusses the use of induced pluripotent stem cells (iPSCs) in various biomedical applications, specifically for toxicity testing. The authors describe their strategies for...

The paper presents a protocol for the large-scale production of renal proximal tubular epithelial cells (PTEC) derived from human pluripotent stem cells (hPSC). The researchers...